Manual No. 085-3435-07
26/45
IGB/IGX Series (Overseas Specifications) Service Manual
2.10 C9 Mode: Board Inspection (A/D Check, I/F Check)
This mode is for factory inspection, and shall not be used for maintenance.
Operation Display
1.
Starting Test Mode
Press
ON/OFF
.
Press and hold
ZERO(+)
while all segments are lit.
2.
Entering C9 Mode
Press
ZERO(+)
eight times.
Or,
press
TARE(-)
.
3.
Press SET.
Span adjusted A/D data is displayed.
4.
Press SET.
A/D converter output value is displayed.
However, the lower 3 bits of 24-bit output change rapidly and
are outside of visual inspection range, so they are excluded.
5.
Press ZERO(+).
Used for A/D board drift inspection.
The inspection result is displayed: OK or Err
6.
Press TARE(-).
Connect the inspection jig and all I/O ports are checked.
The inspection result is displayed: OK or Err
7.
Press COLUMN to return to Key Test Mode.