2.4 Press the ZERO key if the count diverges from 5000
counts.
2.5 Put the weight same as weighing capacity on the
weigh platter, then press the [ * ] key
→
"CAL" is
displayed, then the A/D count becomes "35000" on
the display.
If the count diverges from 5000 counts, unload the
weight and repeat the operations as set out in 2.4 and
2.5.
2.6 When the C3 mode is required, press the TARE key.
- When finishing, press the Memory Switch to
record the data to the memory, and enter the Ending
Test Mode.
•
“CAL” operation can not be performed unless zero-point adjustment is finished.
•
“CAL” operation can not be performed unless the original A/D value exceeds
33300 counts when the weight is loaded.
2.1.9
F Mode – Setting measuring conditions and E2ROM clear
•
All data has been fixed according to country specifications.
•
Altering data may not conform to weighing and measuring tests for a country.
•
See 1.6 “C1 Mode-Country No. Setting” for Measurement Conditions Setup
Table.
1) Operation
Operation
Display
1. Stating
Test
2. Setting measuring conditions
2.1 Press the TARE key for three times
2.2 Press the [ *] key
→
The least significant digit blinks.
Whenever the [ *] key is pressed, the required digit can be
selected.
2.3 Whenever the [ *] key is pressed, the figure of the digit
which has blinked does the increment.