Manual No. 0158A
22/68
IWB
−
IWX Series (Overseas Use) Service Manual
Item 10. C6 Mode
⋅⋅
RAM clear & E2ROM clear
Operation
Display
1. Test Mode start-up
y
Press the [Display ON/OFF] key ON.
y
Keep pressing the [ZERO(+)] key while the all segments
are lighting.
2.
Press the [ZERO(+)] key ON for five times.
3.
Press the [SET] key ON.
4. RAM clear
(Clear the preset data of Registration Mode.)
Press the [ZERO(+)[ key ON.
Press the [ZERO(+)] key ON.
5.
Press the [COLUMN] key ON to return to the Test Mode.
E2ROM clear (Initialization of the E2ROM data)
6.
Press the [SET] key ON.
7.
Press the [Memory Switch] ON.
The Memory Switch: SW1 on the main circuit board.
Note
The C1 data and F01
∼
F17 set data are
initialized. Set up these data again.
8.
Press the [Display ON/OFF] key ON.
After initialized, no information other than C1 is
displayed.
It is necessary to set up the C1 data (Country No., Scale
No., Zero point & Span adjustment).