ITA Series – Table Top Precision Weighing Balance
ITA Series – Table Top Precision Weighing Balance
SYMBOL
NAME
FUNCTION
Power Key
Mode /
Up Shift Key
Tare /
Right Shift
Key
Zero /
Enter Key
10
3
4.CONTROLS & FUNCTIONS
·
·
·
·
·
·
·
·
·
To power on and off the machine
Left Shifts the selected digits during
cal or other setup modes
Switch to the counting mode Switch to Internal Count Mode.
Hold the Mode button and switch on the machine.
Tare the weight kept on the pan
Up shift the selected digits during cal or other setup
modes like 0 to 1 to 2 to 3
Toggles between the options in different setup modes
Zero the weight kept on the pan.
Enter Key – Accept the selected setting
and exit from setting mode.
7.1
INTERNAL USER CALIBRATION
Internal calibration also referred as service calibration requires opening
the machine and changing the jumpers. It is advisable to do this step
only if you have required assistance or the service personnel assisting
you to do this.
Open the battery cover from the bottom
of the machine and remove the cover
and change the calibration jumper
marked as CAL to ON. Make sure there is
nothing kept on the weighing pan.
Turn on the machine and it will read
“CAL”
Press Zero and the display will show “d
1”. This setting is used to set the division
of the scale. Valid values are 1, 2 or 5. You
can change to desired division by
pressing the TARE button.
Press Zero and the scale will now show
the decimal point on the right most digit
of the display. You can set the decimal
point here. Decimal point values can be
set by using the Mode key. The decimal
point will move to the left
STEP
DISPLAY
PROCEDURE
1.
2.
3.
4.