background image

 

 

Printed in Slovenia 

 Subject to change without notice 

 Version 02 / Sep. 2006 

22.600.718

1.

 

Check and correct, if necessary, the zero position of the pointer 
when the instrument is switched off. After cleaning the transparent 
window, eliminate eventual static electrification with a contact or 
a humid mop. 

2.

 

Set the ratings selector switch to the required rating. When 
measuring current and voltages start at the highest rating and then 
change over to the corresponding lower rating. The circuit is not 
interrupted. Be careful about upper limits of ratings as well as 
correct polarity of terminals. After the measurement set the 
selector switch to the highest voltage rating. 

3.

 

Set zero on ohmic ratings at short circuited terminals COM and 50 

µ

A, 100 mV, A, 

4.

 

Consider more detailed instructions in the following chapters for 
meter connection. 

9.

 

MEASURING INSTRUCTIONS 

9.1

 

MEASUREMENT OF DC VOLTAGES 

9.1.1

 

Direct measurement of voltage up to 600 V (20 k

/V) 

 

 
Ratings selector switch: 

600 V … 1 V

DC

 

Reading: a black scale V, A

DC

 

9.2

 

MEASUREMENT OF DC CURRENTS 

Connect the instrument in series with the user into the conductor 
which has lower voltage towards earth. For safety’s sake this voltage 
should not exceed 600 V. 

9.2.1

 

Direct  measurement  of  currents  up  to 1 A and voltage 
100 mV 

 

 
Ratings selector switch: 

1 A … 50 

µ

A, 100 mV 

Reading: a black scale V, A

DC

 

9.3

 

MEASUREMENT OF AC VOLTAGES 

9.3.1

 

Direct measurement of voltages up to 600 V 

 

 
Ratings selector switch: 

600 V … 10 V

AC

 

Reading: a red scale 

9.3.2

 

Measurement of amplification (weakening) in decibels 

The dB scale enables you a direct finding of amplification 
(weakening) at voltage measurement on four-poles. Amplification or 
weakening is proportional to logarithm ratio of voltage square at input 
and output of a four-pole in decibels (dB). The values in dB are 
marked with a positive sign for amplification, and with a negative sign 
for weakening.  

A decibel scale is correspondingly divided into positive and negative 
values from a starting point 0 dB (level 0). The starting point 0 dB is 
defined with 0.775 V voltage according to the international standards. 

Voltages are measured by the described method of measuring AC 
voltages. At 10 V rating the decibels are read directly, while at other 
ratings the rating constant should be added up to the read value by the 
table: 

Rating 

(V) 

10  30 100 300 600 

Constant 

0 +10 +20 +30 +40 

9.4

 

MEASUREMENT OF AC CURRENTS 

Connect the instrument in series with the user always into the 
conductor with lower voltage towards earth. For safety’s sake this 
voltage should not exceed 600 V. 

9.4.1

 

Direct measurement of currents up to 3 A 

 

 
Ratings selector switch: 

3 A … 3 mA

AC

 

Reading: a red scale V, A

AC

 

At measurement at 3 A rating, the measurement should last max. 1 
minute. 
At measurement at 3 mA rating, the circuit resistance should be min. 
150 k

9.4.2

 

Measurement of current by means of a clamp-on meter 

For measurement of AC current without interruption use a clamp-on 
meter. 

 

 
 
Ratings selector switch: 

regarding the ratio 

Reading: red scale V, A

AC

 

 
Consider all instructions for measurement with the clamp-on meter. 

9.5

 

MEASUREMENT OF RESISTANCE 

Periodically check the battery condition and replace a used up battery 
in due time. Before the measurement of resistance always make sure 
that there is no voltage on the measured resistor. Capacitors should be 
discharged. 

At transition to another rating, set the pointer to a final tilt (0 on 

 

scale). When testing the semiconductor, the battery positive pole 
should be on COM socket. 

In 

 x 1 range the battery is extremely loaded, therefore the 

measurement should be as short as posible. 

 

 
Ratings selector switch: 

 x 1 … 

 x 100 

Reading: 

 scale 

 

 
Short circuit COM plug-in sockets and 50 

µ

A, 100 mV, A, 

 and set 

the pointer to the final tilt (0 on 

 scale) with a knob. 

10.

 

MAINTENANCE 

A special maintenance is not required. Periodically check the battery 
condition and voltages. A used up battery should not be left in the 
meter. The battery contacts should be clean and should have a reliable 
contact. If the instrument is not used for a longer period of time, take 
the battery out of the meter. 

10.1

 

REPLACEMENT OF A BATTERY AND A FUSE 

Carefully remove the rubber border. Remove the base by pressing the 
locking ring in the middle of connection sockets on the meter rear side 
with a screwdriver or a similar tool. The meter should be 
DISCONNECTED when removing the rubber border and the base! 
Replace the battery and the fuse. 

10.2

 

CLEANING 

To clean the instrument use a soft mop and a brush. Eventual static 
electrification of the window which could influence in the 
measurement should be removed with an anti-static coating. Remove 
the residue of electrification contact or a humid mop. Use alcohol to 
clean dirt. 

11.

 

REPAIR AND SERVICING 

Please contact an authorised service for repair, servicing and purchase 
of spare parts. 

Reviews: