background image

2-3 Electrical Specifications 

 
Accuracy is ± (% r number of digits) at 23°C ± 5°C, less than 75% R.H. 
 

(1) DC A 

Range 

Resolution 

Accuracy 

200A 

0.1A 

±(1.9%r 5digits) for 0 ~ 20A 

      ±(1.9%r 3digits) for 20A ~ 200A

 

600A 

1A 

     ±(1.9%r 3digits) for 200A ~ 600A 

Summary of Contents for ICM36R

Page 1: ...INSTRUCTION MANUAL ICM36R CLAMP MULTIMETER ICM 36R EN FR IT DE...

Page 2: ...1 ISO TECH ICM36R CLAMP MULTIMETER INSTRUCTION MANUAL...

Page 3: ...2...

Page 4: ...following items 1 Current meter 2 Carrying case 3 Instruction Manual 1 2 Meter Safety Terms as Marked on Equipment Attention Refer to Manual Double Insulation Protection Class II DANGER Risk if electr...

Page 5: ...from slipping off the hand while in use 3 Function Switch Slide switch is used to select the desired range 4 DC Zero Adj Knob Zero adjusts on DC current ranges 5 AC DC Slide Switch Selects a c or d c...

Page 6: ...5 Figure 1...

Page 7: ...reading of 1999 Polarity Automatic polarity indicated Overrange Indication 1 or 1 indicated Measuring Rate 2 times per second nominal Position Error 1 of reading Type of Sensing Hall effect sensing f...

Page 8: ...ient 0 C 50 C 75 R H Storage Temperature 20 C 60 C 80 R H Temperature Coefficient 0 2 x Specified accuracy C 18 C or 28 C Power Requirement Alkaline 9V battery Battery Life 40 hours typical Alkaline M...

Page 9: ...ications Accuracy is reading number of digits at 23 C 5 C less than 75 R H 1 DC A Range Resolution Accuracy 200A 0 1A 1 9 reading 5digits for 0 20A 1 9 reading 3digits for 20A 200A 600A 1A 1 9 reading...

Page 10: ...20A 1 9 reading 5digits for 20A 200A 600A 1A 1 9 reading 5digits for 200A 600A Frequency Response 40Hz 400Hz Data Hold Hold display reading for all functions and ranges 3 Auto Power Off The meter will...

Page 11: ...1 Make sure that the battery is properly connected 2 The instrument should only be operated between 0 C 50 C and at less than 75 R H 3 Do not use or store this instrument in a high temperature or high...

Page 12: ...onductor then take the reading from the digital display 3 3 DC Current Measurement 1 Set the AC DC switch to the DC position 2 Set the function switch to the desired range if unsure select the 600A ra...

Page 13: ...12 Figure 2...

Page 14: ...ly wipe the case with a damp cloth and detergent do not use abrasives or solvents Battery Replacement This instrument is powered by a single 9V alkaline battery Use the following procedure to replace...

Page 15: ...nited Kingdom Japan West Tower 12th Floor Yokohama Business Park 134 Godocho Hodogaya Yokohama Kanagawa 240 0005 Japan USA 7410 Pebble Drive Fort Worth Texas 76118 6961 Canada 1701 Woodward Drive Ste...

Reviews: