IDM 98III / 99III English
IDM 98III / 99III English
20
21
(4) Peak HOLD (For IDM99-III only)
Specified accuracy ±150d.
(5) Resistance
Input Protection
: 1000VDC or 1000VAC rms
Max. Open Circuit Voltage
: Approx. 2.5V for 600Ω & 6kΩ Range.
Approx. 0.6V for others.
Max. Test Current
: Approximately 0.1mA.
* There is a little rolling less than ±50 digits when measuring
>
10.00MΩ.
(6) Continuity Check
Input Protection
: 1000VDC or 1000VAC rms
Max. Open Circuit Voltage
: Approx. 2.5V
Max. Test Current
: Approx. 0.1mA
Continuity Threshold
:
<
30Ω Beep On.
>
100Ω Beep OFF.
Continuity Indicator
: 2.7kHz Tone Buzzer
IDM98-III/IDM99-III
Range
Resolution
Accuracy
600.0Ω
0.1Ω
± (0.8% + 5d)
6.000KΩ
0.001KΩ
± (0.8% + 2d)
60.00KΩ
0.01KΩ
600.0KΩ
0.1KΩ
6.000MΩ
0.001MΩ
40.00MΩ*
0.01MΩ
± (1.0% + 5d)
IDM98-III/IDM99-III
Range
Resolution
Accuracy
600.0Ω
0.1Ω
± (0.8% + 5d)
(7) Diode Test
Input Protection
: 1000V
DC
or 1000V
AC
rms
Max. Open Circuit Voltage
: Approximately. 2.5V
Max.Test Current
: Approximately 0.4mA
(8) Capacitance
Input Protection
: 1000V
DC
or 1000V
AC
rms
Max. Measuring Time
: 0.7 sec for 1nF~1mF
3 sec for 1mF~10mF
(9) Frequency Counter
Input Protection
: 1000V
DC
or 1000V
AC
rms
Min. Frequency
: 1Hz
Sensitivity
:
> 5.0Vp-p (for ACV 1Hz ~ 10kHz).
> 10Vp-p (for ACV 10kHz ~ 100kHz).
>
2mAp-p (for AC mA).
>
0.2Ap-p (for ACA).
IDM98-III/IDM99-III
Range
Resolution
Accuracy
2.000V
1mV
± (1.5% + 2d)
IDM98-III/IDM99-III
Range
Resolution
Accuracy
1.000µF
0.001µF
± (1.2% + 2d)
10.00µF
0.01µF
100.0µF
0.1µF
1.000mF
0.001mF
10.00mF
0.01mF
IDM98-III/IDM99-III
Range
Resolution
Accuracy
100.00Hz
0.01Hz
± (0.1% + 2d)
1000.0Hz
0.1Hz
10.000KHz
0.001KHz
100.00KHz
0.01KHz