Dual Furnace Models M17705, N17706 & 17707 Iss.09 – 12/14 

SPECIFICATION 

 

Dimensions of the cabinet containing the apparatus are: 

 

Width:   

 

600mm 

Depth:   

 

560mm 

Height:  

 

960mm (excluding thermometer storage rack) 

 

Control: 

 

The  heat-pipe  and  the  annealing  furnace  are  controlled  independently  to  a  resolution  of 

0.1°C. 

 
Communications: 

RS422 

 

 

Temperature range: 

120°C to 250°C. 

 

Over-temperature  

Provided on each furnace and operated by an independent sensor. 

protection: 

 

 

 

 

Supply:  

 

2kW, 220-240V, 50/60Hz. 

(110-120V to special order) 

 

Performance:   

Essentially temperature-gradient-free heat-pipes provide the ideal apparatus for fixed point 

calibration and give a virtually perfect freezing profile within the fixed point cell. 

 

The addition of the second furnace enables thermometers to be stabilised before calibration, 

by annealing out strain and removing any 3D oxides. 

 

A comprehensive manual accompanies each apparatus. 

 

Reproducibility of the temperatures of freezing indium or tin is typically ±0.1mK with Isotech fixed point cells. 

 

HOW TO ORDER 

 
Ask for Model 17707.  We need to know your voltage and whether you require communications.  A temperature 

equalising block is an available option (for comparison calibration purposes). 

 

PLEASE  NOTE:  You  may  wish  to  purchase  cells  and  thermometers  with  this apparatus.  Both are available from 

Isotech.  Next Generation Apparatus is also available for Zinc, Aluminium Silver and Copper fixed points.  Please ask 

for data sheets 17705 and 17706. 

 

 

 

 

Summary of Contents for M17705

Page 1: ...9AG England Tel 44 0 1704 543830 Fax 44 0 1704 544799 Internet www isotech co uk E mail info isotech co uk The company is always willing to give technical advice and assistance where appropriate Equa...

Page 2: ...RATION CALIBRATION APPARATUS FOR ZINC ALUMINIUM AND SILVER FIXED POINTS 18 INTRODUCTION 18 DESCRIPTION 18 METHOD OF OPERATION 18 SPECIFICATION 19 THE SODIUM HEAT PIPE FURNACE MODEL 17705 20 FINAL PART...

Page 3: ...COVERED BY THIS GUARANTEE INTERFERENCE WITH OR FAILURE TO PROPERLY MAINTAIN THIS INSTRUMENT MAY INVALIDATE THIS GUARANTEE RECOMMENDATION The life of your ISOTECH Instrument will be prolonged if regula...

Page 4: ...IEC 417 Caution Hot Surface ELECTRICAL SAFETY This equipment must be correctly earthed This equipment is a Class 1 Appliance A protective earth is used to ensure the conductive parts cannot become li...

Page 5: ...is hot or cold unless wearing the appropriate protective clothing and having the necessary training 6 Do not drill modify or otherwise change the shape of the apparatus 7 Do not dismantle the apparatu...

Page 6: ...were carried out Insulation Test M Date Earth Impedance Date Temperature cycle to maximum temperature Date Communications test if applicable Date Laboratory test at C Comments or concessions Controll...

Page 7: ...1 Slacken the central nut using the required tube spanner the insertion of a screwdriver through the tube spanner will prevent the central stud turning 2 Remove the 3 screws securing the transit clam...

Page 8: ...uld be replaced with a fuse A spare fuse is supplied with the furnace FUSE AND NEUTRAL LINK REPLACEMENT Hazardous voltages are exposed when the rear furnace panel is removed Before removing the panel...

Page 9: ...0mm which can be used to pre warm or anneal up to 4 thermometers 3 A pre calibration pocket to install a single thermometer immediately prior to calibration or to house the monitor thermometer between...

Page 10: ...range from near ambient up to 700 C MODEL 17706 Contains a potassium heat pipe with a temperature range from 400 C to 1000 C thus encompassing Zinc Aluminium and Silver freeze points Its annealing fu...

Page 11: ...ate problem or ignored With the dual furnace thermometers can be stored pre warmed calibrated and annealed in one comprehensive facility without introducing contamination or strain The two furnaces in...

Page 12: ...17705 N17706 17707 Iss 09 12 14 ACKNOWLEDGEMENT The Isotech dual furnace owes its origin to Dr P Marcarino of Italy His research and development solved many of the problems of contamination and handli...

Page 13: ...this limit should be exceeded the red push button switch will reset the apparatus once the furnace temperature has fallen below the AL1 value To turn on system power at start up or as required it is n...

Page 14: ...at up rate of the furnace is limited to ensure long life of the heat pipe Impose a heat pipe set point of 150 C and set the over temperature controller to 200 C Allow 1 to 2 hours for the furnace to s...

Page 15: ...By adjusting the set point to a few degrees above the appropriate melting point the cell tin or indium is melted in the essentially temperature gradient free water filled heat pipe When the cell s con...

Page 16: ...ion and give a virtually perfect freezing profile within the fixed point cell The addition of the second furnace enables thermometers to be stabilised before calibration by annealing out strain and re...

Page 17: ...ise and then reduce the over temperature cut out setting until the relay trips thus switching off the furnace Allow the furnace to cool and isolate it from the electricity supply At the rear are two w...

Page 18: ...appropriate melting point the cell zinc aluminium or silver is melted in an essentially temperature gradient free potassium filled heat pipe When melting is complete the heat pipe temperature is read...

Page 19: ...mmunications RS422 Over temperature Provided on each furnace and operated by an independent sensor Protection Temperature range 400 to 1000 C Power 3kW 110V 50 60Hz Performance Essentially temperature...

Page 20: ...ff the furnace Allow the furnace to cool and isolate it from the electricity supply At the rear are two water connections these should be connected to a controllable supply and a flow of 1 to 2 litres...

Page 21: ...his can be accomplished by blowing air at ambient temperature across them by means of a free standing fan 1 By adjusting the set point to a few degrees above the appropriate melting point the cell alu...

Page 22: ...422 Over temperature Provided on each furnace and operated by an independent sensor Protection Temperature range 450 to 1100 C Power 3kW 220 240V 50 60Hz 110 120V to special order Performance Essentia...

Page 23: ...been provided in kit form that can be fitted to dual furnace Models 17705 and 17706 The completed assembly is depicted below ANNEALING COMPONENT OF THE DUAL FURNACE SHOWING AIR FLOW Air percolates thr...

Page 24: ...ure equalising block is for use up to 700 C only Above 700 C the sensors to be calibrated may be immersed directly into the heat pipe without using an equalising block IMPORTANT NOTE Since the calibra...

Page 25: ...t the very low incidence of failure almost implies modes not encountered previously and therefore not easy to envisage before hand Therefore we now prefer to work differently With international commun...

Page 26: ...se A1 935 12 48 A2 Black Fuse Holder 935 11 09 45 amp Opto Relay 935 21 19 110V Contactor 935 21 18 110V Rocker Switch 935 27 02B N Thermocouple 935 14 42 White Fuse Holder 935 11 10 Inconel Cell Bask...

Page 27: ...least 2 hours This operation outgases the basket and insulation which may smoke and discolour during this first temperature excursion The cell can then be inserted into the basket in readiness for use...

Page 28: ...Page 28 of 40 Dual Furnace Models M17705 N17706 17707 Iss 09 12 14...

Page 29: ...Page 29 of 40 Dual Furnace Models M17705 N17706 17707 Iss 09 12 14...

Page 30: ...Page 30 of 40 Dual Furnace Models M17705 N17706 17707 Iss 09 12 14...

Page 31: ...is to contain the appropriate material in a sealed enclosure from which all other materials including air have been evacuated leaving a space to be filled by the vapour phase at a pressure appropriat...

Page 32: ...gon In practice the difficulties of setting up conditions to facilitate this measurement can conveniently be circumvented by carrying out the alternative procedure of comparison calibration in which t...

Page 33: ...3 ITL M 17668 ITL M 17701 3 4 7 TIN FP 231 928 505 078 3 3 2 2 ITL M 17669 ITL M 17701 3 4 7 ZINC FP 419 527 692 677 4 3 2 7 ITL M 17671 ITL M 17701 2 3 6 ALUMINIUM FP 660 323 933 473 7 0 1 6 ITL M 1...

Page 34: ...au temperature The cell is a fragile device Although it is as rugged as is consistent with its materials and purpose it must still be regarded as a kilogram or more of mass loosely contained in a fran...

Page 35: ...can no longer be assumed to be gas tight It may leak its enclosed gas and atmospheric air may leak into it The pressure at the freeze point may as a consequence be incorrect and more seriously contam...

Page 36: ...However the technique requiring association of cells with other equipment involves subtlety and operator sensitivity Before relying upon measurements made in them the operator should perform enough fr...

Page 37: ...IONAL SERVICES AND INFORMATION Isotech operates one of the world s most comprehensive UKAS supervised Laboratories Training is available to customers at an agreed daily rate The Isothermal Journal of...

Page 38: ...Page 38 of 40 Dual Furnace Models M17705 N17706 17707 Iss 09 12 14 FIGURE 1 SECTION THROUGH DUAL FURNACE...

Page 39: ...ss 09 12 14 FIGURE 2 SEALED METAL FREEZING POINT CELL A High purity graphite crucible and cover B High purity graphite sleeve C High purity metal D Fused quartz envelope filled to give a pressure of 1...

Page 40: ...Page 40 of 40 Dual Furnace Models M17705 N17706 17707 Iss 09 12 14 FIGURE 3 SEALED METAL FREEZING POINT CELL BODY DIMENSIONS...

Reviews: