©Isothermal Technology
Page 15 of 61
914 millisKanner manual Iss. 1.00
link cable terminated with a Lemo connector and a universal power supply. Additional
millisKanners are provided with analogue link cables terminated with stackable 4mm
plugs at both ends and a DC power link cable so that you can create the required
expanded system.
First connect a single millisKanner to your milliK (see section 2.2 above). Use the null
modem cable provided (see section 8.1 if you wish to make your own cable) to connect
the unused RS232 port on the rear of additional millisKanners to the unused RS232 port
on the previous millisKanner to form a ‘daisy-chain’ arrangement. The RS232 ports on
the millisKanner are completely interchangeable, so it does not matter which is used to
connect to the next millisKanner or the milliK in the system. Similarly, use the Power Link
cable provided to connect either DC Power connector on the millisKanner to the unused
DC Power connector on the previous millisKanner in a ‘daisy-chain’ arrangement.
Connect the five Output Terminals on the additional millisKanners (in parallel) to the first
millisKanner (the analogue cables provided use plugs that allow “piggy-back” connection
to make this easy):
If you wish to make your own analogue link cables, use a screened cable with low
dielectric loss insulation such as PTFE or polyethylene (otherwise measurement
performance may be affected). The screen should be connected to the ground terminals
on both millisKanners.