I
V
The connection between the positive (red) terminals should carry no current and should
be connected as close to the base of the I+ terminal as possible in order for the 3-wire
compensation to work effectively. The lowest 3-wire compensation error is achieved by
using a wire under the screw terminal to connect the positive terminal together and then
adding the 'single' PRT lead to the binding post:
The top (green) terminal should be connected to the screen of the PRT’s lead (where
present) to minimise electrical noise picked up by the wires. The thermometer type in the
setting screen must be set to “Default 3-Wire PRT” or a user-defined thermometer that
has been set to 3-wire in order for the 3-wire compensation to occur.
3.4
3.4
3.4
3.4
Connecting PRTs (2
Connecting PRTs (2
Connecting PRTs (2
Connecting PRTs (2-
-
-
-wire)
wire)
wire)
wire)
The millisKanner can be connected to 2-wire PRTs. The two red and two black
measurement terminals should be connected together and then connected to the PRT as
follows: