©Isothermal Technology
Page 26 of 61
914 millisKanner manual Iss. 1.00
R
B
R
B
G
-
+
Reference
Junction
Junction
Cu
Cu
M+
M -
Isothermal Environment
RED
BLACK
RED
BLACK
GREEN
I
V
R
B
R
B
G
+tc
This technique is less accurate than using an ice-point due to the additional
uncertainties associated with the PRT or thermistor used to monitor the reference
junction and any temperature gradients between the temperature of the PRT/thermistor
and the thermocouple’s reference junction.
3.5.3
3.5.3
3.5.3
3.5.3
Connecting Thermocouples (
Connecting Thermocouples (
Connecting Thermocouples (
Connecting Thermocouples (using
using
using
using internal compensation)
internal compensation)
internal compensation)
internal compensation)
To use a thermocouple with internal reference junction compensation (reference junction
formed within the millisKanner and its temperature measured using an internal sensor),
fit a thermocouple connector of the same type as the thermocouple and connect this to
the socket below the binding posts. In the ‘Settings’ screen for the thermocouple
channel, set the ‘Reference Junction’ to “Internal”: