©Isothermal Technology
Page 52 of 61
914 millisKanner manual Iss. 1.00
Deselects (disconnects) all inputs (on the millisKanner to which the RS232 connection is
made).
Example: to deselect all inputs use:
OPEN
8.2.6.8
8.2.6.8
8.2.6.8
8.2.6.8
Command: MI
Command: MI
Command: MI
Command: MILLi
LLi
LLi
LLiskanner:LIST?
skanner:LIST?
skanner:LIST?
skanner:LIST?
Format: MILLiskanner:LIST?
This command is used by the milliK to ‘discover’ which millisKanners are connected in
the ‘daisy-chain’. It reports information on the millisKanner and also forwards the
command to the next millisKanner in the daisy chain (on the alternate RS232 port) so
that it will also respond.
Example: if there are two millisKanners (serial numbers 13-P543 and 13-P544, firmware
1.00) connected in the daisy chain with a milliK (serial number 13-P432, firmware 2.00)
at the other end of the chain, the response to
MILL:LIST?
will be:
Isothermal Technology,millisKanner,13-P543,1.00
Isothermal Technology,millisKanner,13-P544,1.00
Isothermal Technology,milliK,13-P432,2.00
8.2.6.9
8.2.6.9
8.2.6.9
8.2.6.9
Command: MI
Command: MI
Command: MI
Command: MILLi
LLi
LLi
LLiskanner:INITialise
skanner:INITialise
skanner:INITialise
skanner:INITialise
Format: MILLiskanner:INITialise
<start>
This command is reserved for use by the milliK and should not be used directly. It is used
to assign channel numbers to each millisKanner in the daisy-chain. It assigns channel
number
<start>
to the first input and sends the same command with
<start>
incremented by 10 to the next millisKanner in the daisy-chain (on the alternate RS232
port).