©Isothermal Technology
Page 54 of 61
914 millisKanner manual Iss. 1.00
Example: to deselect channel 17, use:
MICR:OPEN17
8.2.6.13
8.2.6.13
8.2.6.13
8.2.6.13
Command: MI
Command: MI
Command: MI
Command: MILLi
LLi
LLi
LLiskanner:
skanner:
skanner:
skanner:RJC:
RJC:
RJC:
RJC:OFFSet
OFFSet
OFFSet
OFFSet#
#
#
#????
Format: MILLiskanner:RJC:OFFSet<channel#>?
Reports the calibrated RJC sensor offset (in °C) for the specified channel.
Example: a millisKanner (assigned channel numbers 20-29) with an RJC sensor offset of
0.123°C will respond to
MILL:RJC:OFFS25?
with “0.123”.
8.2.6.14
8.2.6.14
8.2.6.14
8.2.6.14
Command: MILLiskanne
Command: MILLiskanne
Command: MILLiskanne
Command: MILLiskanner:RJC:T
r:RJC:T
r:RJC:T
r:RJC:TEMP
EMP
EMP
EMPerature#?
erature#?
erature#?
erature#?
Format: MILLiskanner:RJC:TEMPerature<channel#>?
Reports the temperature of the reference junction (in °C) for the specified channel.
Example: a millisKanner (assigned channel numbers 20-29) with a reference junction
temperature of 20.123°C will respond to
MILL:RJC:TEMP25?
with “20.123”.
8.2.6.15
8.2.6.15
8.2.6.15
8.2.6.15
Command: MI
Command: MI
Command: MI
Command: MILLi
LLi
LLi
LLiskanner:RST
skanner:RST
skanner:RST
skanner:RST
Format: Command: MILLiskanner:RST
This command is reserved for use by the milliK and should not be used directly. It is
used to reset all the millisKanners in the daisy-chain by performing a reset (equivalent to
a power-on reset) and passing the command to the next millisKanner in the daisy-chain.
Example: on power-up a milliK will send the command
MILL:RST
to reset all the
millisKanners connected to its RS232 port.
8.2.6.16
8.2.6.16
8.2.6.16
8.2.6.16
Command:
Command:
Command:
Command: MEA
MEA
MEA
MEASure
Sure
Sure
Sure::::RJC?
RJC?
RJC?
RJC?
Format: MEASure:RJC?