Page 17 of 49
TTI-22 Iss.04– 11/15
7. Configuring TTI-22
After being switched on, TTI-22 is in measuring mode. Press the <M> key to change to the menu mode.
There are two ways to select a menu item:
•
Use the arrow keys to move the cursor to the desired item and press the <
> key.
•
Enter the digit in front of the menu item.
If the display does not provide enough space for all lines in a menu, the symbol for the arrow key <
> or
<
> appears on the right-hand side in the headline. You can scroll down the display with these arrow
keys.
To go back to the next higher level, press the <ESC> key.
To change back to the measuring mode, press the <ESC> key in the main menu.
7.1
Setting a Password
TTI-22 has been assigned no password ex factory. We recommend setting a user password as soon as
possible.
The user password restricts the access to the menus Edit Configuration, Change User Password and
LOCK/UNLOCK Device.
To prevent unauthorized access to the menu, change to measuring mode before leaving the instrument or
switch TTI-22 off.
If you have forgotten your user password, contact your Isotech representative.
Setting a user password
Fig. 7 - 1
Setting or changing the user password
1.
Choose the menu 3 Change User Password.
2.
Enter the old user password ("Old Password").
When setting a password for the first time, enter "0000" in this line and confirm with<
>.
3.
Enter a new user password ("New Password").
4.
To save the new password, press the <
> key and select the answer "1 YES" to the question "Save
changes?".