15
9.3
HOW TO USE EARSCOPE
This medical device is used in conjunction with a speculum (sizes 3, 4, 5mm) which allows
you to perform visual examinations of the hearing canal outside of the walls of the eardrum
through direct analysis of otoscopic photos or digital video caught by the device.
The speculum is made of reusable polypropylene.
In this configuration the device offers doctors a valuable method for improving the
accuracy of the visual analysis.
The measurements are made by DinoCapture software previously installed on a personal
computer.
To be used as otoscope the Dino-Lite requires the deployment of a speculum CE marked
made of reusable polypropylene. The speculum requires a special procedure for cleaning
and disinfection as described in the special paragraph.
Put the speculum on the device before capturing pictures or video (simple joint) and ask
the patient to take a comfortable position and to remain immobile for the duration of the
visit.
•
Connect the microscope to the computer
•
Start the software DinoCapture (previously installed)
•
Moisten the contact cap and the skin of the patient with disinfectant or oil for
dermatoscopy
•
Put the contact cap on the skin of the patient
•
Choose image or video to capture
•
Zoom acting on the wheel to obtain the required size of the picture see
previous page amendments
•
Verify on the focus wheel the current magnification
•
Captures pictures or videos using microtouch
•
Move the device from the patient
•
Place the device on the Amita support (if present)
•
Tell the patient is able to move again
•
Move your mouse cursor on the picture preview just took the photo on the left
side of the screen (Picture 1)
•
On the top of the screen you can enter patient information and other notes
deemed necessary (Picture 2)
•
Click on Save as choosing the desired resolution (Picture 3)
•
Save in your computer using the coding system of the patient deemed
necessary (Picture 2)
•
To make measurements see Chapter 10
Earscopic investigations must be achieved by keeping the device parallel to the
floor of the room in order to minimize the risk of eardrum perforation caused by
the fall of the microscope or speculum
The device is intended for external use only.
Measurement or visual surveys of the ear and the hearing canal must be
conducted exclusively by using a speculum (CE marked)
Make sure that the Dino-lite has been regularly connected to the PC prior to the
application of the speculum on the body of the device