22
13.
Enter the unit of measure by selecting the “Unit” dropdown box.
14.
Press “Finish” when done.
Hint: You can create multiple calibration profiles for different Dino-lite models.
The calibration can be done through a small or large value of magnification, the
most accurate calibration will be achieved when the zoom value is closer to that
used for the measure
The calibration can be removed by selecting "Clear Calibration" menu "ruler"
The measures to be addressed without deleting the calibration disabling
"Measure with Adjusted Mag."
"The Adjusted Mag.”, will not be shown if the calibration is off
WARNING! USE A METROLOGICAL GRADUATED SCALE of 1 OR 2
MILLIMETERS TO MAKE CALIBRATION
12 FREQUENCY OF CALIBRATION
Calibration has to be made:
•
At the first use
•
Each time the device is used after leaving it idle for at least 10 days
whenever the measurements taken give improbable or contradictory results