Mic-Fi Wi-Fi Microscope User Manual
page 25
7.2
Calibration Menu
No Calibration: Use the default measurement settings.
New Calibration: Create a new calibration profile.
Open Calibration: Opens the folder where your calibration setting are located. From that window, you can select
already created calibration profiles or delete them.
It is strongly suggestion: To obtain the accurate measurement data, please calibrate before any measurement
function at each magnification.
Select a known measurement standard as the to-be-observed object, such as a ruler, to be used for calibration. Bring
photo of the selected ruler into focus by turning the “Focus & Magnification adjustment wheel”, determine the
magnification by taking a reading from the scale on the focus dial as below(The default magnification is 60):
1.
In the live video window select the calibration drop down menu.
2.
Select “New Calibration”.
3.
A small window will pop-up, please enter profile name
、
magnification and the actual length.
4.
Drag the mouse to draw a line which has actual length.