22
Password:
OFF: password disabled
Use +/- to set a password first digit, press ? to
go to the next digit.
Confirm with ENT.
Level or flow sensor. +/- to select LEVEL to connect
a level sensor, or for a flow sensor.
Confirm with ENT.
Time to exit the menu after the last button pressed.
+/- to change value.
Confirm with ENT.
Delay time to activate the pump.
+/- to change value.
Confirm with ENT.
To solve problems of inertia of some electrodes, or installations, the delay time
switch on of the pump can be set.
CALIBRATION
press ENT to calibrate
pH calibration:
POINT 1.
Put the sensor in buffer solution pH7.
+/-, to modify value on the display to get 7.00.
Confirm with ENT.
POINT 2
Wash the sensor with water and put it in the second
buffer solution (pH 4 or pH9).
+/- to modify value on the display to get the value of
the buffer solution.
Confirm with ENT.
INPUT MODE
LEVEL
MENU
SETUP
MENU
SETUP
MENU
CALIB
DELAY
h:m:s
DELAY
h:m:s
STARTUP
MENU
CALIB
CALIB
pH
2
POINT
CALIB
pH
1
POINT