23
ORP (redox) calibration:
POINT 1: Short circuit the BNC connector.
+/- to select 0.
Confirm with ENT.
Point 2: Connect the sensor and put it in the chemical solution (250, 475, 650 mV).
+/- to select the same value of the chemical solution used.
Confirm with ENT
SET POINT
ENT to set the setpoint.
+/- to set the set point value.
Confirm with ENT
Select dosing method. If the pump has to dosify to
increase the reading to reach the set point or
decrease the reading. To dosify acid, select , to
dosify alkaline product .
+/- to select. Confirm with ENT.
HYSTERESIS (only in FULL menu) , is the diference
from the set point where the pump starts or stops.
+/- to change value.
Confirm with ENT.
Type of control.
ON-OFF: to get the set point, the pump works at a
fixed frecuency.
+/- to select. Confirm with ENT.
PROP: Frecuency of the dosing depends on the
reading and the set point distance.
+/- to select. Confirm with ENT.
MENU
SETPOINT
SET
pH
POINT
SET
DIRECTION
POINT
HYST
POINT
SET
pH
MODE
SET
ON-OFF
POINT
MODE
SET
PROP.
POINT