Operating manual uni-cart, vexio-cart, pro-cart, duo-cart, compact-cart, classic-cart and endo-cart 09/2019
9
IMPORTANT INFORMATION
been tested and approved by appropriate, qualified personnel!
•
Ensure the isolating transformer is only connected to a power supply with a functioning
protective earth connection that complies with the provisions of the IEC 60364-7-710
2012-10 “Electrical installations in buildings, part 7-710: Requirements for special
installations or locations – Medical locations”. If in doubt, contact a specialist in the
electrical trade or an authorised employee of the hospital’s engineering team.
•
Personnel (hospital and service personnel) working directly or indirectly with a mobile
equipment carrier must be instructed!
•
Setting adjustments may only be carried out by qualified personnel.
•
Repairs may only be carried out by ITD GmbH.
Safe working with the equipment cart:
• The device is only disconnected from the mains supply when the power plug is removed
from the socket.
Operation:
•
With every change of location it must be ensured that no one is injured or objects damaged!
Connections:
•
In the USA and Canada, use a hospital-grade connection cable when connecting isolating
transformers in the 115 V mode and in Japan use the optional Japanese connection cable.
•
When connecting to the multiple-socket strip, secure the plug using an optional available
plug pull-out lock.
•
Only devices that fulfil the requirements of IEC 60601-1 or are IEC-tested may be connected
to the sockets/connecting lines.
•
Additional medical equipment with connecting bolts for equipotential bonding are to be
connected with the green-yellow cable to the optional equipotential bonding connecting bolts!
Caution: The overall rating given on the nameplate must not be exceeded.
Please observe that no further multiple sockets must be connected to an
existing multiple socket.
Load capacity:
•
The total weight of the equipment and the accessories on the mobile equipment carrier
may not exceed the permitted payload weight (see load capacity sticker on the base
frame).
•
The surface load imprinted on the system components may not be exceeded!
•
The load shown on the fittings (e.g. infusion stand, jointed arms) may not be exceeded!
Infection protection:
•
Hygiene regulations are to be observed when cleaning!
•
Give only cleaned and disinfected equipment and fittings to a service technician for
maintenance and repair work!
Environmental protection:
•
Dispose of all cleaning and disinfection agent residue in a manner not harmful to the
environment!
2 Assembly
2.1 Completeness
Unpack the uni-cart, vexio-cart, pro-cart, duo-cart, compact-cart, classic-cart or endo-cart
EN
Summary of Contents for aCart Compact
Page 3: ......
Page 5: ......
Page 31: ...EN ...
Page 57: ...ES ...
Page 83: ...За верността на съдържанието отговаря ITD GmbH BG ...
Page 109: ...CZ ...
Page 135: ...DA ...
Page 161: ...DE ...
Page 187: ...NL ...
Page 213: ...26 EST ...
Page 239: ...FI ...
Page 265: ...FR ...
Page 317: ...HU ...
Page 343: ...IT ...
Page 369: ...HR ...
Page 395: ...LAT ...
Page 421: ...LIT ...
Page 447: ...NO ...
Page 499: ...PT ...
Page 525: ...RO ...
Page 551: ...SK ...
Page 577: ...SW ...
Page 603: ...SV ...