16
Please note:
Make sure that this is also the case for adaptation of equipment in the power supply circuit
(e. g. multiple socket strip).
3.9 EMC
Electromagnetic compatibility of the medical electrical equipment positioned on the cart has
to be checked by the overall system configurer. If you use a different equipment combination,
you should check electromagnetic compatibility between individual items of equipment before
using in a medical setting.
Customer-specific equipment trollies used within the nuclear spin environment must be tested
by the customer for suitability for use due to the ferromagnetic materials they contain. ITD GmbH
excludes any liability in this respect.
Using the carrier system with an isolating transformer within an environment subject to explosive
hazard is not permissible.
3.10 Excluded from the final electrical inspection of system components
and
accessories
ITD GmbH exclude the following system components and accessories from the final electrical
inspection:
•
Multiple socket strips without additional protective conductors that are not wired in the
mounting
•
ME cables and appliance cables included
•
POAG plates and cables included
ELECTRICAL SAFETY
EN
Summary of Contents for aCart Compact
Page 3: ......
Page 5: ......
Page 31: ...EN ...
Page 57: ...ES ...
Page 83: ...За верността на съдържанието отговаря ITD GmbH BG ...
Page 109: ...CZ ...
Page 135: ...DA ...
Page 161: ...DE ...
Page 187: ...NL ...
Page 213: ...26 EST ...
Page 239: ...FI ...
Page 265: ...FR ...
Page 317: ...HU ...
Page 343: ...IT ...
Page 369: ...HR ...
Page 395: ...LAT ...
Page 421: ...LIT ...
Page 447: ...NO ...
Page 499: ...PT ...
Page 525: ...RO ...
Page 551: ...SK ...
Page 577: ...SW ...
Page 603: ...SV ...