SYSTem Subsystem
SYSTem:POSetup <CPD>
This command is used to set some parameter settings or working status when
the instrument is powered on.
•
RST: The parameter setting or status when the instrument is powered on is
equivalent to executing the
*RST
command.
For details, see
.
•
SAV0: The parameter setting or status when the instrument is powered on is
equivalent to executing the
*SAV 0
command.
For details, see
Group
System
Syntax
SYSTem:POSetup <CPD>
Arguments
<CPD>
RST|SAV0
Default Value
RST
Returns
None
Example
SYST:POS SAV0