Ti300 PRO, Ti400 PRO, Ti401 PRO Ti450 PRO, Ti450 SF6, Ti480 PRO
Users Manual
14
Level/Span
Level and Span are values within the total range of temperature set in
Range
. Level is the temperature level to
view within the total range of temperatures. Span is the span of temperatures to view within the total range of
temperatures. See Table
In automatic Level/Span mode, the Imager sets
Level/Span
based on the temperatures set in
Range
.
When the
Range
of the Imager is set to one of the preset measurement ranges and
Level/Span
is set to
Manual
,
the level setting moves the thermal span up or down within the total temperature range.
To change the Level/Span:
1. Select
Measurement
>
Level/Span
>
Manual
.
2. Select
Set Level/Span
.
Table 9. Level and Span Settings
Item
Description
Level
Span
Total Imager range
1
2
3
www.