Warranty
HI144
logger and communication cradle are warranted for
two years
against defects in workmanship and materials when
used for their intended purpose and maintained according to
instructions.
This warranty is limited to repair or replacement free
of charge. Damages due to accident, misuse, tampering or lack of
prescribed maintenance are not covered. If service is required,
contact your local Hanna Instruments Office. If under warranty,
report the model number, date of purchase, serial number
and the nature of the problem. If the repair is not covered by
the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first
obtain a Returned Goods Authorization (RGA) number from the
Technical Service department and then send it with shipping
costs prepaid. When shipping any instrument, make sure it is
properly packaged for complete protection.
All rights are reserved. Reproduction in whole or in part is
prohibited without the written consent of the copyright owner.
INSTRUCTION MANUAL
HI144 T-Logger
Hanna Instruments reserves the right to modify the design, construction or appearance
of its products without advance notice.
Operational Guide
Technical Service
This instrument is factory calibrated, but it is generally
recommended to have all thermometers recalibrated at least
once a year.
For an accurate recalibration, contact your nearest Hanna
Customer Service Center.
Initial Operation
The logger is supplied complete with a CR2032 lithium battery.
Open the battery compartment cover (see “Battery Replacement”
section for details), insert the battery with the + sign facing
up and tighten back the compartment cover.
The display will show the
“PC”
message
to indicate that the logger needs to be
programmed through a PC.
Startup
In order to communicate with the logger, the application
software must be installed on your PC.
Connect
HI144002
communication cradle to your computer
through a USB port, and place the
HI144
on the cradle.
Run the application software. The yellow LED on the cradle
blinks to indicate that the communication is active.
Do not
remove the logger while the LED is blinking
.
The following operating parameters can be set:
• Measurement Unit: °C or °F
• Starting Type: PC Start, Button Start or Date/Time Start
(hh:mm:ss)
• Logging Method: Wrap Around Memory, until memory is full (at
8000 samples) or number of samples (a maximum of 8000)
• Logging Interval: from 1 minute to 24 hours (1 minute
resolution)
• High and Low Alarm thresholds
• Logger Password
Click on the
Program
key to program the logger with the
selected options.
www.
.com
information@itm.com
1.800.561.8187