SETUP PROCEDURE
SETUP PROCEDURE
5
1. Carefully unpack the TURBO TEMP™ Electronic oral and
rectal probes, and carrying strap, checking to ensure that
each item is undamaged.
2. Using the adhesive mounting tape attached to the back of
the dispenser, or the screws supplied, attach the Probe Cover
Dispenser Bracket (Model 896) to a convenient location and
install a box of 200 probe covers.
3. Set the instrument to the °C (Celsius) or °F (Fahrenheit) scale,
as desired, by adjusting the recessed switch on the left side of
the battery compartment with a small screwdriver or pen tip
(°C=right switch position; °F=left switch position) see
Figure 1a.
4. Attach the home base to a convenient location using either
the adhesive tape or screws supplied with the instrument. The
home base may be attached to either a horizontal or vertical
surface.
5. Install 3 size AA batteries and install cover. *
6. Insert the connectors on the probe cords (oral or rectal) into
the connector sockets on the back of the instrument, and
route the probe cord through the cord guide (see Figure 1a).
Insert the probes into the probe storage wells in the top of
the instrument. Insure the oral (blue) probe plug is
connected to the #2 socket and the rectal (red) probe is
connected to socket #1.
*
Remove batteries if thermometer is not likely to be used for some time.
INST
ALLA
TION
Installation
NOTE: Batteries must be
installed exactly as shown on
the battery compartment or
the instrument will not
function. Install negative end
of battery first to prevent
damage to the battery
contacts. To open battery
compartment, press one side
of battery cover until latch
releases, then press other
side (refer to Figure 1a).
Dispose of exhausted or used
batteries in accordance with
applicable regulations.
NOTE: Whenever batteries
are replaced, the instrument
will perform a self-test
routine of the instrument
hardware which will take
about 15 seconds. Should
the instrument fail a self-
test, a FIX ME message
along with a code number
will appear in the display.
Refer to the Calibration Self-
Test and Service Information
sections of this manual for
further information. At the
end of the self-test routine,
the instrument will emit
beeps continuously and
display
RETURN TO
B A S E
. The TURBO TEMP™
Electronic Thermometer
must be returned to the
home base to reset the anti-
theft timer whenever
batteries are replaced.
*
*
Danger: The Turbo Temp
™
Electronic Thermometer is an
explosion hazard, do not use in
the presence of flammable
anesthetics.
*