Getting Started
9 / 120
5001FAOPT71 ISS 8.0
Purge
1.
PURGE
can only be operated with the cover
opened and the key switch in the
RUN
position.
2.
Press the
PURGE
buttons together until fluid
flows and priming of the syringe extension line is
complete. The audible alarm will operate during
use of the
PURGE
buttons and the volume used
during priming will be shown in the volume
infused display.
The
PURGE
button allows the delivery of a limited volume of fluid in order to purge the extension line prior to
being connected to a patient.
Important
:
The purge feature is not activated when the cover is closed. Ensure that the extension line
is disconnected from the patient before purging the line. No alarms are disabled during the
operation of the purge feature.
Front Panel and Main Display
Time
Drug Infused
Pump Status
Demand
Status
Current Protocol & Syringe Type
Clinician Over-ride
The clinician over-ride feature can be used in
RUN
mode to administer an additional bolus dose or continuous
background infusion of a limited dose and duration, for example during the PCA lock out period. The clinician
over-ride is a special feature which can be configured according to the specific clinical situation. Clinician over-
ride can also be used in
SET
mode to allow modification of the pre-set PCA Protocol when this option has been
disabled for normal use.
1.
Turn key to RUN position and ensure green light is
illuminated on the START button.
2.
Press and hold down the
CLINICIAN OVER-RIDE
button for 2 seconds.
3.
Use “+” and “-” arrows and
NEXT
button to enter
three figure pre-programmed clinician access code
“n n n”. See technical service manual.
4.
Select
BOLUS
or
CONTINUOUS.
5.
Use “+” and “-” arrows to select the dose delivered,
when the correct value has been entered press
OK.
6.
Use “+” and “-” arrows to select period over which
the dose is to be delivered. Press
OK
when correct
time has been entered.
7.
BEGIN BOLUS? YES
- Clinician bolus /
continuous infusion will be delivered to the patient.
NO
- Quit set up and return to normal operation.
Important:
The delivery of the clinician over-ride continuous infusion will automatically halt while a
Patient or Clinician over-ride bolus is being administrated. To cancel clinician over-ride
during delivery, press STOP and press the YES softkey.
Pumping
Pressure
Maximum
Dose