Getting Started
14 / 120
5001FAOPT71 ISS 8.0
When ever a new protocol is selected, modified or the key switch is turned to the
SET
position,
the
CONFIRM PROTOCOL
display will appear. Pressing
OK
automatically records the protocol
and any changes in the
EVENT LOG
history.
Using Pre-Set PCAM Protocols
Operation of PCAM is greatly simplified by the use of PRE-SET PCA protocols. When the PCAM
key switch is turned to the
SET
position the unit will automatically display pre-set
PROTOCOL A
if
NEW PATIENT
has been selected, or, display the previous protocol in use if NEW PATIENT has
not been selected.
With the key switch in the
SET
position, it is possible for the user to modify the pre-set protocol
using the
MODIFY PROTOCOL
button and select another pre-set protocol using the
NEXT
PROTOCOL
button.
To modify a Pre-Set Protocol
1. Press
MODIFY PROTOCOL
indicated on
the screen.
2. The protocol summary will be displayed.
Use
ARROWS
to highlight a field, press
ALTER
to enter that field and “+”, “-”
arrows to select desired values.
3. When field is correct press
CONFIRM
or
CANCEL
.
4. Display will show:
5. Press
OK
to return to display.
(
Note:
A modified protocol has no pre-set
letter).
Protocol
Drug name
Drug ‘X ’
Drug conc.
1.0mg/ml
PCA Dose
1.0mg
(1.0ml)
Lockout period
5 mins
Occlusion level
4
Continuous
0µg/h
(0.0ml/h)
Loading dose
0µg
(0.0ml)
Max. limit
50mg
(50.0ml)
Limit duration
4 hours
PCA delivery
STAT
A
To change to another pre-set protocol
Press
NEXT PROTOCOL
to display the step through the pre-set protocol.
Pre-set protocols are identified as A to E. When the desired protocol has been selected it can be
used by turning the key switch to the
RUN
position, or, can be modified using the
MODIFY
PROTOCOL
.
To confirm protocol
Drug ‘X’