1. ICONS on display
YES - displays Pumping Pressure and Maximum Dose ICONS.
NO - ICONS disabled.
2. Protocols in use
PCA pre-set protocols to be available. Select number from 1 to 5.
3. Modify protocol
YES - allows protocols to be modified in SET mode.
NO - removes modify protocol option in SET mode.
4. Handset mode
MODE
A
B
C
BEEP
GOOD
ALL
ALL
HAND SET LIGHT:
PCAM STOPPED
OFF
OFF
OFF
PCA AVAILABLE
ON
ON
ON
PCA DELIVERING
FLASH
ON
FLASH
PCA LOCK-OUT
OFF
ON
ON
5. Delayed call-back
YES - call-back alarm can be delayed from 10 mins up to 90 mins.
NO - call-back will be cancelled for up to 2 mins or extended to 15 mins. To
extend call-back alarms, press and hold the stop key for 4 seconds .
This allows the time to be extended.
6. Display Sleep
YES - display goes blank after 2 minutes.
NO - display stays on during operation.
7. Chirp low alarms
YES - "chirp" alarm during use on battery/near end.
NO - no "chirp" alarm.
8. Continuous infusions
YES - Continuous infusions option in protocols.
NO - Continuous infusions are not available.
9. Loading doses
YES - Loading dose option appears in protocols. To activate this option "NEW
PATIENT" is confirmed. The protocol also includes the loading dose. Start the
PCA.
NO - Loading doses are not available.
10. Max. dose limits
YES - Dose limit option appears in protocols.
NO - Dose limits are not available.
11. Variable dose rates
YES - Variable dose rate option available.
NO - Loading doses are not available.
12. Comms identity number
Use arrows to set pump identity (between 000 and 127) for use with remote
communications.
13. Comms enabled
YES - RS232 Communications enabled.
NO - RS232 Communications disabled.
14. Nurse call
YES - Nurse call connector enabled.
NO - Nurse call connector disabled.
15. Nurse call inverted
YES - Nurse call inverted enabled.
NO - Nurse call inverted disabled
16. Continuous Print
YES - Allows printing of events as they happen.
NO - Continuous printing disabled.
17. Default Syringe
YES - Default syringe enabled.
NO - Default syringe disabled
18. Lock syringe type
YES - Syringe type locked to default syringe.
NO - Syringe type not locked to default syringe.
19. Quiet mode
YES - Quiet mode enabled.
NO - Quiet mode disabled.
Configured Options
20 / 120
5001FAOPT71 ISS 8.0
The general options available are as follows:
General Options