Configured Options
22 / 120
5001FAOPT71 ISS 8.0
Pre-Set Protocol Set Up
1. Turn key switch to
SET
position while
pressing down
START
button.
2. Enter access code using rate selection
arrows.
3. Enter first digit and press
NEXT
indicated
on the screen.
4. When the complete code has been entered
press
ENTER.
5
Select
PROTOCOL DEFAULT SET-UP
from the menu. Press
ENTER.
6. Press
MODIFY PROTOCOL
to display
current protocol summary. Use
ARROWS
to highlight a field, press
ALTER
to enter
that field and "+" and "-" arrows to select
desired values.
7. When field is correct press
CONFIRM
or
CANCEL.
Press
OK
to return to protocol
summary.
Pre-set Drug names and safety limits apply to pre-set protocols during protocol selection and
modification.
Note:
One of the five drugs programmed in the Drug Names and Safety Limits are
selectable for each protocol. Additionally a default drug with limits set to
extremes may be chosen and is indicated by “ DRUG “.