Specifications
25 / 120
5001FAOPT71 ISS 8.0
Self Test Sequence
INDICATOR LED TEST:
Check that the
STOP
and
START
LED’s are
flashing.
DECLUTCH SWITCH TEST:
Squeeze plunger holder finger grips and
check that the display alternates between 1 (engaged) and 0
(disengaged - finger grips squeezed together).
NEAR END OF INFUSION TEST:
Starting with an empty, extended
50/60ml syringe on the unit, squeeze the finger grips and move the
plunger holder slowly to the left.
Check the display switches from 0 to 1. The position at which the display
changes will depend on the syringe type being used. ( For example 6ml
on a B-D Plastipak 50/60ml syringe and 3ml on a Terumo 50ml syringe -
see Technical Service Manual for values).
LINEAR GRID TEST:
Squeeze the finger levers and position the drive
mechanism to the right then slowly move the syringe plunger to the left
observing the display alternate between 0 (OPTO over slot) and 1 (OPTO
over bar).
PLUNGER DETECTOR TEST:
Press plunger plate button. Check the
display switches from 1 ( No syringe fitted) to 0 ( Syringe plunger fitted).
MOTOR ENCODER TEST:
Motor is pulsed while encoders are tested.
COVER DETECT TEST:
Open and close the cover. Check that display
changes from 0 (cover open) to 1 (cover closed).
SYRINGE SIZE DETECTION SYSTEM:
Lift syringe clamp and check
that the number shown in the display increases within the normal range
(approx. 045 to 215).
PUMPING PRESSURE DETECTION TEST:
Remove syringe and
confirm that value is within normal range (-020 to +020). Gently press
back on the plunger holder and observe the value increase.
PATIENT HAND SET TEST:
Connect patient hand set. Check display
switches from OFF to ON as the button is pressed.
KEY SWITCH TEST:
Turn key from
SET
position to
RUN
position. Check
display changes.
Summary of user defined options selected.
Returns unit to normal operation.
7.
LED’s Flashing
8.
DECLUTCH
9.
EOI opto:0
10.
Grid opto :0
11.
Plunger opto:1
12.
Motor/encoder:1
13.
Cover detect:0
18.
Nursecall on/off/on....
14.
Syringe pot:_ _ _
15.
Beam value:_ _ _
16.
Handset: OFF
17.
Key switch : SET
19.
CONFIGURATION
SUMMARY
20.
NEXT
Summary of Contents for P5000 Mk II
Page 2: ......