Specifications
32 / 120
5001FAOPT71 ISS 8.0
Routine Maintenance Procedures
To ensure that this instrument remains in good operating condition, it is important to keep it clean
and carry out the routine maintenance procedures described below. All servicing should only be
performed by a qualified service engineer with reference to the technical service manual (TSM)
for this product (Service Manual reference: 5000PB00004).
Refer to the Technical Service Manual for user access codes.
INTERVAL
ROUTINE MAINTENANCE PROCEDURE
As Required
Thoroughly clean external surfaces of the pump before and after
prolonged period of storage.
12 Monthly
1. Inspect AC power supply plug and cable for damage.
2. Perform functional OPERATIONAL AND ALARM TEST. See
TSM.
3. Perform the SELF TEST ROUTINE.
4. Perform rate accuracy verification test. See TSM.
5. Perform pressure calibration verification test. See TSM.
6. Operate the pump on battery power until the battery low alarm
then charge the battery to confirm battery operation and
charging.
Important:
If the pump is dropped, damaged, subjected to excessive moisture or high
temperature, immediately take it out of service for examination by a qualified
service engineer.
Disposal
The pump should be disposed of taking environmental factors into consideration. To ensure no
risk or hazard remove the internal rechargeable battery and the nicad battery (component B1)
from the control board and dispose of as outlined by the local country regulations. Do not send
back to manufacturer. All other components can be safely disposed of in the normal manner.
Cleaning and Storage
Before the transfer of the pump to a new patient and periodically during the use, clean the pump
by wiping over with a lint-free cloth lightly dampened with warm water and a standard disinfectant
/ detergent solution.
The syringe and extension line are disposable single use items and should be discarded after use
according to their manufacturers instructions.
If the pump is to be stored for an extended period it should be first cleaned and the internal battery
fully charged. Store in a clean, dry atmosphere at room temperature and, if available, employ the
original packaging for protection.
Once every 3 months during storage, carry out functional tests as described in the
OPERATIONAL AND ALARM TEST in the technical service manual and SELF TEST ROUTINE
and ensure that the internal battery is fully charged.
Important:
Before cleaning always switch OFF and disconnect from the AC power supply.
Never allow liquid to enter the casing and avoid excess fluid build up on the
pump. Do not use aggressive cleaning agents as these may damage the exterior
surface of the pump. Do not steam autoclave, ethylene oxide sterilise or immerse
this pump in any fluid.