Specifications
34 / 120
5001FAOPT71 ISS 8.0
Occlusion Pressure Limits for IVAC 50ml Syringes
The following tables show the worst case values for line pressure, time to alarm and bolus volume
that can be expected in the event of an occlusion when the IVAC 50ml syringe is selected,
G40020 administration set.
Alarm Level
Rate (ml/h)
Maximum Time
Nominal Occlusion
Maximum Infusion
Maximum Bolus
to occlusion alarm
Alarm Pressure
Pressure
Volume
(min:sec)
(mmHg)
(mmHg)
(ml)
0
1.0
2:00
0
50
0.1
1
1.0
8:00
92
110
0.2
2
1.0
20:00
184
220
0.3
3
1.0
33:00
276
330
0.5
4
1.0
52:00
368
450
0.7
5
1.0
65:00
460
560
0.9
6
1.0
85:00
552
670
1.0
7
1.0
102:00
664
780
1.2
8
1.0
120:00
736
80
1.6
9
1.0
140:00
828
1000
1.8
10
1.0
155:00
920
1100
2.0
0
5.0
01:00
0
50
0.1
1
5.0
02:00
92
110
0.2
2
5.0
05:00
184
220
0.3
3
5.0
07:00
276
330
0.5
4
5.0
10:00
368
450
0.7
5
5.0
12:00
460
560
0.9
6
5.0
15:00
552
670
1.0
7
5.0
17:00
644
780
1.2
8
5.0
20:00
736
890
1.6
9
5.0
24:00
828
1000
1.8
10
5.0
26:00
920
1100
2.0