Specifications
35 / 120
5001FAOPT71 ISS 8.0
Spare Parts
A comprehensive list of spare parts for IVAC Syringe pumps is included within the service manual.
This can be ordered from ALARIS Medical Systems, or authorised distributor. For part number
please refer to summary parts list below:
Part Number
Description
5001FAOPT71
PCAM Directions for Use P5000 Mk II GB/FR/D.
5000PB0004
PCAM Service Manual. Model P5000.
5000SP00008
Printer Cable 9 to 25 pin (9 - pin female (PCAM) to 25 - pin
male (printer)).
5000SP00010
Citizen PN60 Printer Cable (9 - pin female (PCAM) to 26 - pin
AMP17823404).
1000SP01008
Comms Cable (9 - pin female to 9 - pin female)
0000EL00004
Internal Battery - 6v NP2.6-6 Rechargeable.
0000EL00287
Fuse - T63mA (Time Lag 5 x 20mm) - 220/240V units
0000EL00280
Fuse - T12mA (Time Lag 5 x 20mm) - 110/120V units
1001FAOPT91
AC Power Cable - UK
1001FAOPT92
AC Power Cable - European
1001FAOPT93
AC Power Cable - Unterminated
0000ME00026
Foot - Self Adhesive
1000SP01015
Pole Clamp Assembly
5000LB00020
Label Set P5000
5000LB000023
Label - Front Panel
Bolus Volume Accuracy
The following table provides an indication of the accuracy with which a bolus infusion will be
delivered. Test carried out as specified in IEC601-2
Bolus Volume
Bolus Rate
No. of Samples
Max. Positive
Max. Negative
Mean
(ml)
(ml/hr)
(%)
(%)
(ml)
0.1
100
25
+12.0
-14.0
-5.0
2.0
100
25
+2.5
-0.0
+1.0
5.0
100
25
+1.0
-0.0
+0.8
Summary of Contents for P5000 Mk II