background image

This quick start guide only provides general operating instructions. For more detailed 
information, you can download the AFC V1 User Manual at 

support.izon.com

INTENDED USE

The AFC is used to automatically collect fractions from qEV columns. The instrument is 
intended for use in research laboratories by professional personnel for research use only. For 
verification of the entire system, it is recommended that Good Laboratory Practices (GLPs) be 
followed to ensure reliable analyses.

COMPONENTS OF THE AFC SYSTEM:

1. 

AFC Instrument with attached  
Nozzle Set and Silicone Waste Tube

2. 

Nozzle Set (x3)

3. 

qEV Column Mount (x5)

4. 

Carousel Cover

5. 

Reversible Carousel

6. 

Carousel Plate

7. 

qEVoriginal Reservoir

QUICK START GUIDE

AUTOMATIC FRACTION COLLECTOR V1 

(AFC)

8. 

5 mL Collection Tube (x15)

9. 

2 mL Collection Tube (x30)

10. 

12 Volt Power Supply

11. 

Power Lead

12. 

USB 2.0 Type A to USB 2.0 Micro Cable 
(male to male)

13. 

10 g Calibration Weight

The carousel is reversible and can be used in two orientations: 

 

One side has 

seven large holes

 and uses 5 mL collection tubes.

 

One side has 

thirteen small holes

 and uses either 1.5 mL or 2.0 mL collection tubes. 

Cannot be used with the qEV10.

 

qEV100 is not compatible with the AFC.  

To complete a run on the AFC you will also need:

A qEV column, pipettes and tips, freshly prepared and 0.22 μm filtered and degassed buffer, 
freshly prepared 0.5 M NaOH solution for column cleaning, 70% ethanol solution for cleaning 
and a computer with Windows 10 for updating the AFC operating firmware as required. 

1

www.izon.com

Reviews: