During calibration, the Agilent power meter will display the momentary frequency and level of the
output signal being measured. The frequency will vary from 2MHz to 1640MHz (see
Figure 4.8
)
Please note that the Power meter has been turned into remote mode by the FireFox (
RMT
is displayed
in the LCD) and the unit is communicating as evidenced by the
RX/TX
indicators. The Power meter
may sometimes also display the
ERR
indicator, this indicator can be disregarded.
Please note that the power meter indicates that it is using sensor calibration table 2 with the
(2)
indication in the LCD display under the frequency. You must enter and enable the calibration table
properly to avoid calibration errors.
Figure 4.8 Agilent power meter display during calibration