FireFox Synthesizer User Manual
62
© 2006 Jackson-Labs
The unit will display a message as shown in
Figure 4.9
when the calibration is finished successfully.
Calibration data is only stored in Flash memory at the end of the calibration when this message is
displayed.
N.B.: TO USE THE FIREFOX, YOU MUST UNPLUG THE UNITS’ POWER AFTER AND
ONLY AFTER THE CALIBRATION IS FINISHED AND YOU SEE THE MESSAGE IN
Figure 4.9
You can verify proper operation of the FireFox RF output using the power meter as it is connected for
calibration. The FireFox output power from -30dBm to +5dBm should be -1dBm, and are
typically -0.2dBm after calibration when measured with the setup describe above, and if the
power meter is set to the proper frequency using its
Frequency
button.
Figure 4.9 Successful Calibration results