FireFox Synthesizer User Manual
1.3 General
1.3.1 Environmental Conditions
This instrument is intended for indoor use. It is designed to operate at a maximum relative
non-condensing humidity of 95% and at altitudes of up to 4000 meters. Refer to the specifications
tables for the ac mains voltage requirements and ambient operating temperature range.
1.3.2 Line Power and Grounding
The FireFox Desktop unit operates on a 100V to 220V, 43Hz to 65Hz AC power input with typically
less than 20W power consumption. The FireFox OEM unit operates on a 12.0V regulated DC input
with typically less than 1A input current. Both units are designed to be continuously powered to keep
the GPS receiver and OCXO warmed-up and locked.
The FireFox desktop unit requires a three-pronged AC power plug with proper grounding to prevent
electrocution hazards. All BNC and SMA connector shields are grounded, and should not be
connected to any voltage sources. To minimize shock hazard, the instrument chassis and cabinet must
be connected to an electrical ground. The instrument must be connected to the ac power supply mains
through a three-conductor power cable, with the third wire firmly connected to an electrical ground
(safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or
disconnection of the protective earth terminal will cause a potential shock hazard that could result in
personal injury. If the instrument is to be energized via an external autotransformer for voltage
reduction, be certain that the autotransformer common terminal is connected to the neutral (earthed
pole) of the ac power lines (supply mains).
1.3.3 Line Fuse
Fuses are internal to the instrument and are not user-serviceable. All repairs should be referred to a
qualified service technician. The FireFox OEM module uses a self-resettable Polyfuse that will
correct itself once the fault or overload condition is removed.
1.3.4 Do not operate the Instrument in an Explosive
Atmosphere
Do not operate the instrument in the presence of flammable gases or fumes. The instrument contains
mechanical switches and relays that could cause sparks and thus ignition of flammable gases or
fumes.
1.3.5 Keep away from Live Circuits
Operating personnel must not remove instrument covers. Component replacement and internal
adjustments must be made by qualified service personnel. Do not replace components with power
cable connected. Under certain conditions, dangerous voltages may exist even with the power cable
Summary of Contents for FireFox
Page 1: ...FireFox Synthesizer User Manual Document 0012008 Version 004 Date 10 April 2008...
Page 2: ...FireFox Synthesizer User Manual Copyright 2008 Jackson Labs...
Page 34: ...FireFox Synthesizer User Manual 28 2006 Jackson Labs...
Page 46: ...FireFox Synthesizer User Manual 40 2008 Jackson Labs SOURce POWer...