Setup Commands are used to specify instrument and measurement parameters for the next
measurement. To specify more than one parameter, sequential setup commands may be sent to the
instrument before the measurement is initiated.
The default parameters are those used during the measurement immediately before Remote Mode
operations are initiated. If Remote Mode operations are under way, the default setup values for the
upcoming measuring are those defined in the last Setup (S_ command).
Setup
Command
Description
SA
Select Add-on Accessory 1
An Add-on accessory is one that is used in conjunction with a primary accessory. For
example, a neutral density filter (Add-on Accessory) used with the MS-75 (Primary
Accessory). Up to 3 Add-on accessories can be specified for a measurement.
Syntax
: SAn[CR]
Where
: n = Accessory code
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: Accessory Codes can be found by running report 116 (command D116). See the Data
Codes section for specific details.
SB
Select Add-on Accessory 2
An Add-on accessory is one that is used in conjunction with a primary accessory. For
example, a neutral density filter (Add-on Accessory) used with the MS-75 (Primary
Accessory). Up to 3 Add-on accessories can be specified for a measurement.
Syntax
: SBn[CR]
Where
: n = Accessory code
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: Accessory Codes can be found by running report 116 (command D116). See the Data
Codes section for specific details.
SC
Select Add-on Accessory 3
An Add-on accessory is one that is used in conjunction with a primary accessory. For
example, a neutral density filter (Add-on Accessory) used with the MS-75 (Primary
Accessory). Up to 3 Add-on accessories can be specified for a measurement.
Syntax
: SCn[CR]
Where
: n = Accessory code
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: Accessory Codes can be found by running report 116 (command D116). See the Data
Codes section for specific details.