99
PR-805/810/810
L
User Manual
M
EASUREMENT AND
D
ATA
S
END
C
ODES
Measurement and Data Send Codes are used to measure (
M
Command) and then specify returned data
or acquire values without making a measurement (
D
command).
Either a
D
or an
M
can precede any of the following codes. For example,
M1 or D1
–
M602 or D602
. If an
M command is sent, a measurement will always be made even if the Data Code does not request
measured values. For example, if
M116
is sent to the instrument, a measurement is made then a list of
accessories is returned from the instrument.
In the following table,
<STAT>
is the returned error code. If
<STAT>
is all zeros (00000) no error has
occurred during the request. All other values for
<STAT>
relate to an error condition. Refer to the
Remote Mode Error Code section of the manual for a complete list of error codes and their meanings.
<Units>
in the output format is the photometric unit type of the measurement per the following table:
Type
Code
Units
Luminance
111
fL
cd/m²
Illuminance
112
fc
lux
Luminous Intensity
113
mcd
Luminous Flux
114
lumens
TABLE 10 - PHOTOMETRIC UNITS CODES
Note:
Both <STAT> and <Units> are contained in every output and are not annotated below.
All data fields are fixed length (except where otherwise noted) and comma delimited. Commas also
serve as placeholders for empty fields.