  WEIGHING SCALE

 

D. Wind Shield 

 

Note: Always reinstall the transportation protection device before transporting the scale, lest the precision 
sensor should be damaged due to bumping and falling during transportation. 

 

IV. PRECAUTIONS 

 

1. 

Full charge the battery after unpacking the scale.

 

Recharge the battery

: When battery symbol appears on the LCD display, charge the battery with AC power cord 

plug in, the indicator of charge will light up in red; when it becomes green means charge completed. (It takes 
about 8 hours to full charge the battery.)

 

2.  Install the equipment on a level and stable surface. 
3.    Do not install the equipment near the air conditioning or a vibrating machine. 
4.  Install the equipment in an environment with steady temperature (0

~40

), avoid rapid 

temperature changes. 

5.  Independent AC outlet for this equipment is recommended, check the voltage before plug in. 
6.  Warm up the equipment for 15 minutes before use. 
 

V. KEYPAD 

 

1.

PRINT

 

Press PRINT key, RS-232 will print data, when negative value is shown on the screen, the 

printing function is prohibited. 

 

2.

MODE

 

Press MODE key for selecting weighing unit. 
 

3.

TARE

 

Press TARE key for deduction of current weight on the weighing pan. After removing the weight 

on the weighing pan, press TARE key to cancel the original tare. 

Note: This operation can only be performed in the stable mode. 
 

4.

ZERO

 

Press ZERO key and the screen will show zero 

0.0000

, press ZERO key in the tare mode, the tare 

will be cancelled; if the weight on the weighing pan is within the zero adjustment range, the 
scale will be reset to zero. 

 

5. “ON/OFF” 

Power switch. 

Summary of Contents for LPWN-1530

Page 1: ...electronic hardware plastic medicine textile and various other industries It is useful for packaging inventory and various production and quality control cases It will help you save your time labor mo...

Page 2: ...haped iron piece inside hole 7 Then fix the three hole aluminum column 6 and scale support 5 on the shaped iron piece of the scale by tightening the M5 Allen screw 4 using an Allen wrench 3 Put the st...

Page 3: ...he equipment in an environment with steady temperature 0 40 avoid rapid temperature changes 5 Independent AC outlet for this equipment is recommended check the voltage before plug in 6 Warm up the equ...

Page 4: ...y Yes No Page 5 Item 11 1 Power on while holding down MODE key The scale will enter into function setting mode and CAL appears on the LCD display Press PRINT key the LCD display will show the scale mo...

Page 5: ...gain the LCD display will show RS232 which means the scale enters into printing mode setting mode press TARE to select Prt Co Prt St or Prt Pr Pr means data sent when key pressed St means data sent au...

Page 6: ...y shows CAL 2 Press TARE key to enter into single point calibration mode the display will show On0 wait until CAP appears 3 Press MODE key to select calibration weight the display will show in turn CA...

Page 7: ...attery symbol Charge the battery with power on Operating at normal weighing mode Err5 Checking if the weighing object is over the max capacity 9e Calibrating Err6 Change weight APPENDIX 2 RS 232 OUTPU...

Page 8: ...12oz T W 0lb 0 64oz G W 2lb 8 76oz 5 Jin G N W C J i n g C R L F weight Example N W 2 275jin T W 0 037jin G W 2 312jin 6 HK jin G N W _ G t l H C R L F weight Example N W 1 14 08G tl H T W 0 0 48G tl...

Page 9: ...M WEIGHING SCALE Example Total 0 pcs 9 C R L F Example 0 G GROSS N NET T TARE Printer and output format example kg SH 24 EZ2 S N W 1 6025 kg T W 0 0000 kg G W 1 6025 kg N W 1 1375 kg T W 0 0180 kg G...

Reviews: