SW-4010Q-MCL-M52 User Manual (Tentative)
Preparation
- 32 -
SWIR
1. Select
User
to save the gain correction value with
SWIRPixelGainCorrect
.
Note:
You cannot perform calibration when
Off
or
Factory
is selected.
2. Gain correction data is automatically generated by
SWIRPixelGainCalibration
and saved
in the user area specified in step 1.
3. You can check the execution result of gain correction by
SWIRPixelGainCalibrationResult
.
For more information, see the following Short ASCII command table.
Item
Short ASCII
Command
Values
Command
Examples
SWIRPixelGainCorrect
IRPGC
0:Off (Default)
1:Factory
2:User
IRPGC=
[Param.]<CR><LF>
IRPGC?<CR><LF>
Note:
Factory stores
correction data with
factory settings.
SWIRPixelGainCalibration
IRPGR
-
IRPGR<CR><LF>
SWIRPixelGainCalibrationResult IRPGS
0: Pixel black correction has not been finished
yet.
1: Succeeded.
2: Error1 –Image was too bright.
3: Error2 –Image was too dark.
4: Error3 –Timeout occurred.
IRPGS?<CR><LF>