Xonic
®
100 Series
Jain Technology
Tel. 82-2-856-4114 Fax. 82-2-856-9503
7
Step 3:
Adjusting of sensor
direction
The Open Channel transducer angle is fixed to 60°. The angle is measured from the
CENTER of the sensor to the mounting wall.
In order to receive a good signal for best accuracy, all transducers must face diagonally.
For trapezoidal or circular channels, rotate the sensors on both walls so that they face
each other.
Step 4:
Install Flowmeter
and Connect
Cables are integrated with transducers, so connect the other side of cable to the flow
computer.
Summary of Contents for Xonic 100 Series
Page 43: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 43...
Page 44: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 44...
Page 45: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 45...
Page 46: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 46...
Page 47: ...Xonic 100 Series Jain Technology www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503 47...