4. Path Angle: the angle between path length and path distance (CH width).
For LTR-S, LTR-M, LTR-L sensors, the angle is fixed to 60°.
5. Path Height Option: Increase the measuring range for each path.
6. Auto Install: Do not use Auto Install for Open Channel. It is suggested to
use Reserve Install for a better performance.
7. Reserve Install: *click SET for each path to enable the auto installation.
(After setup the information for Path One. User needs to go back to 2. Path
Select to select other paths and setup the install parameters
individually in
4. Install.)
8. Install Info: User can review the install information of the selected path.
9. Actual Zero: User should perform Actual Zero when the flow is actual
stopped but the flowmeter shows other values instead of “0”.
10. Sampling Clk: Flow computer will select Sampling Clock automatically,
therefore, user does not need to change.