Page 36
Problem
Possible Cause
Corrective Action
“Service” Indicator On.
Low salt level in pool water.
Test salinity. Add salt as described in
Section 4.7
.
Cell requires cleaning.
Refer to Maintenance Procedure for acid wash
cleaning. See
Section 6.4, Step 2
.
Cell not working properly after cleaning
Contact your authorized service representative.
DC power cord not properly connected to the cell.
Check connection. Check for dirt or corrosion
around DC cord contacts at Cell. Check Cell studs
for same.
Flow/Temp/Salinity Sensor not plugged in.
Plug Flow/Temp/Salinity Sensor into connector on
power center printed circuit board. (Cycle Power).
Water temperature too low
Winterize your pool. See
Section 6.6, Winterizing
.
All the above have been checked and indicator light is
still on.
Contact your authorized service representative.
Problem
Possible Cause
Corrective Action
Salt Level Too Low.
Not enough salt added on start up.
Add salt to pool, 3.0 gpl to 3.5 gpl.
See
Section 4.7
.
Heavy Rainfall.
Add salt to pool, 3.0 gpl to 3.5 gpl.
See
Section 4.7
.
Leak in pool.
Repair pool.
Dirty Flow/Temp/Salinity Sensor contacts
or faulty Flow Sensor.
Remove and clean Flow/Temp/Salinity Sensor
contacts. Check Flow/Temp/Salinity Sensor cable
for damage. See
Section 6.5
. Verify salinity with
Myron “L” meter calibrated for NaCl, titration test
units, or other salt meters. Please be aware that salt
test strips have a large variance in readings
(400 - 800 ppm).