User Manual Mobile Power System
Pagina 11
5
Inspection and maintenance
The Mobile Power System is a low maintenance system. Maintenance is
therefore limited to inspecting and/or measuring of following points:
1.
Check on possible damage of the housing, check weldings and
screws. Check if possible damage of the housing can course
internal (electrical) hazards.
2.
Visual inspection of wiring and electrical components.
3.
Checking for wear and tear of mechanical parts (including the fan
and switch). Replace when necessary.
4.
Check the condition of the battery. If the battery time is too short for
the application, they should be replaced. Always replace batteries
by the same type.
5.
Verification of correct operation:
a.
Check for correct switching between mains voltage and
voltage generated by the inverter.
b.
Check the output voltage.
c.
Checking the LED indicators and the battery discharge
indicator.
6.
Measure the earth leakage current on both primary and secondary
side.
7.
Measure the earth resistance of grounding and equipotential earth.
Jansen Medicars cannot be held responsible for damage, consequential
damages and injuries caused by incorrect use/treatment or by negligent
maintenance.