7
- Where the unit is exposed to dust.
- Where the unit is exposed to water vapor.
- Where the unit is exposed to salty atmosphere.
- Where the unit is exposed to explosive gas.
- Where the unit is exposed to excessive shocks or vibrations.
- Where the angle of inclination of mounting surface exceeds 10 degrees.
- Where the unit is exposed to direct sunlight.
8) This equipment has been tested and found to comply with the limits for medical devices
to the IEC 60601-1-2:2007. These limits are designed to provide reasonable protection
against harmful interference in a typical medical installation. This equipment
generates uses and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to other devices in
the vicinity. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to other
devices, which can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the other
device(s) are connected.
- Consult the manufacturer or field service technician for help.
9) Do not to touch signal input, signal output or other connectors, and the patient
simultaneously.
10) a statement that MEDICAL ELECTRICAL EQUIPMENT needs special precautions
regarding EMC and needs to be installed and put into service according to the EMC
information provided in the ACCOMPANYING DOCUMENTS;
11) a statement that portable and mobile RF communications equipment can affect
MEDICAL ELECTRICAL EQUIPMENT.
12) Please consult a physician or a trained health professional for interpretation of
measurement results.
Caution
1. Measurements may be impaired if this device is used near televisions,
microwave ovens, X-ray equipment or other devices with strong electrical
fields. To prevent such interference, use the meter at a sufficient distance
from such devices or turn them off.
2. Incorrect operation or failure of user to maintain the unit spares the
manufacturer or his agent of the responsibility for system‟s non-compliance
with specifications or responsibility for any damage or injury.