P/N 960-000180R_Rev. 1 {EDP #233427}
© 2016, JAPAN CASH MACHINE CO., LTD.
6 - 9
Calibration and Testing
DBV® Series DBV-400 Banknote Validator
Section 6
11. Once the Serial Number Setting is completed,
Calibration Data will be written on the EEPROM.
When the data writing finishes normally, the
“
Calibration Succeeded.
” message dialogue box
appears. Click the “
OK
”
button (Figure 6-42
a
) to close the message dialogue box.
12. Click the “
Start
”
button (Figure 6-43
a
) to save
the calibration settings.
13. When the “
Write EEPROM Succeeded.
” screen
appears, click the “
OK
”
button (Figure 6-44
a
).
14. Click the Close
button (Figure 6-43
b
) to
shut down the Calibration application.
This completes the Sensor Calibration procedures.
Figure 6-42
Figure 6-42
Calibration Succeeded Message
a
Figure 6-43
Figure 6-43
Calibration Result
a
b
Figure 6-44
Figure 6-44
Write EEPROM Succeeded
a