Chapter 9
Technical specifications
Characteristics of the options
User Manual
790000002/00
73
Vibration
The 2000 Platform resists the following vibration tests:
–
Complete test comprising 6 cycles along each of the x, y and z axes.
–
One cycle of 5 to 200 Hz and back to 5 Hz with a sweep duration of
one minute/octave.
–
3 mm amplitude displacement test, for the range 5 Hz to 15 Hz.
–
3g acceleration test for the range 16 Hz to 200 Hz.
Flammability
The 2000 Platform housing (in ABS, type V0) does not propagate fire.
Characteristics of the options
Power meter
option
Specifications given for 25°C, after 20 minutes stabilization time and after
zero setting.
–
Wavelength range: 800 to 1650 nm in steps of 1 nm
–
Calibrated wavelengths: 850 / 1310 / 1490 / 1550 / 1625 nm
–
Accuracy at calibrated wavelengths: ± 0.2 dB (at -30 dBm)
–
Input power range : -60 dBm to +10 dBm
–
Maximum resolution: 0.01 dB / 0.01nW
–
Measurement range:+5 to -50 dBm (-45 dBm from 800 to 1250 nm)
–
Linearity within the measurement range: ± 0.2 dB
VFL
–
Wavelength: 650 nm
–
Length of fiber: up to 5 km
–
Class 2 laser (standards EN60825-1 and FDA21 CFR Part
1040.10).
Bluetooth
Option
–
Bluetooth Option
–
Class 2
–
Range: up to 10 meters
2000 Platform English rev00.book Page 73 Lundi, 20. juin 2011 1:47 13