GSM/GPRS/EDGE SIGNAL ANALYZER
34
JD7105B Base Station Analyzer
[Stop Limit]
Enter a value
[dBm]
[Measure Type]
<Examine>
or
<Full>
Press the
Stop Limit
soft key.
Use the numeric keys.
Press the unit soft key to complete the input.
Press the
Measure Type
soft key to toggle between
Examine
and
Full
.
GSM Spurious Emissions Measurement Procedure
MEASURE SETUP
Sets the measurement parameters including range table and measurement type. The measurement
settings can be saved and recalled as a file. Also, JDViewer, PC application software allows the user
to set the parameters in the user’s computer to import into the instrument.
RANGE TABLE:
Sets the range table parameters.
RANGE:
Adds a new range number or selects one of the existing range numbers to show or
hide the particular frequency range in the range table at the bottom of the screen. Use the
numeric keys to enter a range number and press the Enter soft key to complete the input.
Each time the Range soft key is press, it toggles between On and Off.
START FREQUENCY:
Sets or changes the start frequency value to be measured for the
selected range number in the Range menu. Use the numeric keys to enter a value and press
one of the unit soft keys.
STOP FREQUENCY:
Sets or changes the stop frequency value to be measured for the
selected range number in the Range menu. Use the numeric keys to enter a value and press
one of the unit soft keys.
RBW:
Sets the resolution bandwidth for the selected range number in the Range menu.
VBW:
Sets the video bandwidth for the selected range number in the Range menu.
START LIMIT:
Set the start value of the limit line for the displayed range.
STOP LIMIT:
Set the stop value of the limit line for the displayed range.
MEASURE TYPE:
Sets the measurement type to
Examine
or
Full
. Toggle between
Examine
and
Full
. Examine mode will show selected range only. The instrument will automatically switch and
show all ranges when Full is selected.
RANGE:
Selects the range number to be displayed on the screen. Use the numeric keys to select
one of the existing range numbers that are configured in the Range Table menu. Pressing the
Enter soft key completes the input and changes the display.
SAVE/RECALL SETTING:
Saves the measurement parameter settings as a file or retrieves a
measurement setting file saved in the instrument or generated by JDViewer software application.