49
GSM/GPRS/EDGE SIGNAL ANALYZER
JD7105B Base Station Analyzer
AMP/SCALE
REFERENCE LEVEL:
Sets the maximum Y-axis range when using the Spectrum Analyzer or
Signal Analyzer function.
ATTENUATION
AUTO:
Input attenuator’s value is automatically set depending on the reference level. This is
the default operation mode, so it is recommended to use this mode unless a special
measurement condition is needed.
MANUAL:
Sets the attenuator’s value for the
Spectrum Analyzer
RF In 50
Ω
port. The
reference level changes according to the change of the attenuator’s value, but the
attenuator’s value does not change when the reference level is changed.
!
If the input signal level is higher than +30 dBm or higher than the allowable mixer
input limit per input attenuator’s value set by the user, a “IF Overload” message will
be displayed on the screen indicating that the input attenuation value must be
increased. When the instrument is exposed to an over power condition for a long
time, its performance may be degraded.
PREAMP:
Turns on or off the internal pre-amplifier. The pre-amplifier is to amplify a low-level
input signal. This soft key toggles between the
On
and
Off
. When it is turned on, the allowable
attenuation input range is 0 dB to 10 dB with 5 dB step. If user sets attenuation value beyond 10
dB, then the instrument will automatically turn off pre-amplifier to display low level signal properly
on the screen.
EXTERNAL OFFSET:
Sets an external offset value. An offset consists of a cable loss and a user
offset and the measurement result shows the value reflecting both offset values. When the
external offset value is set at 40 dB in the Spectrum mode, the measurement result compensates
40 dB at both the Spectrum Analyzer and Signal Analyzer modes.
Setting Averaging
Step
Description
1.
Set the averaging
BW/AVG
[Average]
Press the
BW/AVG
hard key.
Press the
Average
soft key.
2.
Enter number of averaging to set
Use the numeric keys and input a number between 1 and 99.
When the averaging reaches to the number set here, a new
measurement value replaces the earliest measurement value
in sequence.